Squirrel Input Type : Thermocouple K / T

· Temperature Measurement using Thermocouples

Thermocouples consist of a thermoelement, that is a junction of two specified dissimilar wires that poduce a small temperature dependent voltage.

  • K Inputs are intended for use with Type K thermocouples, that is Chromel/Alumel (Ni-Cu/Ni-Al) thermocouples. Temperatures up to 1370ºC can be measured. Type K sensors are preferred and should be used unless there is a very good reason not to!
  • T Inputs are intended for use with Type T thermocouples, that is Copper/Constantan (Cu/Cu-Ni) thermocouples, permitting temperature measurement up to 350ºC with greater accuracy than Type K thermocouples.
  • CJC (Cold Junction Compensation) is built into the Squirrel (for explanation of CJC see FAQs). This, together with thermocouple connectors and special connecting leads available for thermocouple sensors assures reliable and accurate temperature measurement.
  • Refer to the table "Special Range Setting" for inputs assigned to other thermocouple types such as C, D, E, J, N, R, S and enhanced K and T.

Ranges available in the Squirrel data logger:

Input Type Min Max Units Resolution
K -200 200 °C 0.1
K -200 450 °C 0.2
K -200 1200 °C 0.5
T -200 200 °C 0.1
T -200 350 °C 0.2
Note: Squirrel can display °C or °F on request

Squirrel modules available:

Module Thermocouple inputs Other inputs
M03 4 4 Voltage/Current
M05 8 -
M25 16 -

Note: Other ranges are available to order e.g. -100 to 100°C Type T

 


Home | Corporate | News | Products | Support | Services | Agents | Contact