Squirrel Input Type : P

· Temperature Measurement using PRTs

P Inputs are for use with Pt100, Platinum Resistance Thermometers (PRTs).

  • PRTs can provide high accuracy over a wide range of temperatures.
  • As the Pt100 change in resistance is very small per ºC, special consideration must be given to the interconnecting cables and method of measurement. To minimise the effects of resistance and electrical pickup or interference special connecting cables must be used.
  • The Squirrel supports 2, 3 or 4 wire sensors. However for best results 4 wire sensors should be used.

Ranges available in the Squirrel data logger:

Input Type Min Max Units Resolution
P -200 100 °C 0.1
P -200 400 °C 0.2
P -200 600 °C 0.3

Squirrel modules available:

Module Pt100 inputs Other inputs
M06 4 4 Voltage/Current
M04 8 -

 